Thickness Logging Compliant with AlphaGage+

Wear rates can be calculated and used to estimate when the minimum allowable thickness will remain with the AlphaGage+


The AlphaGage+ CPDL measures the wall thickness of a pipe or vessel and hence can be used to identify and measure corrosion and/or erosion of the material. Wear rates can be calculated and used to estimate when the minimum allowable thickness will remain. The instrument can be used for mapping of pipes and vessels with the result held in a table (T-log) for storage, analysis and reporting.

 

 

Thickness Logging - AlphaGage+
Fig. 1: AlphaGage+

 


 

Recommended Package

Category Part Number Description
Product ALPHAGAGE+ CPDL Corrosion gauge with accurate thickness data logging
Probe PDEM3750 Dual element compression transducer
Frequency: 5 MHz
Diameter: 3/8"
 
Other probes are available upon request.
Contouring probes for pipes are available.
Cable 152062
or
152060
Twin transducer cable TPC-LD
or
Twin transducer cable TPC-BD

 

Frequency recommendations according to the thickness (T):

T > 50mm = 2.25MHz
15mm < T < 30mm = 5MHz
2mm < T < 15mm = 10MHz

For further information, please contact the Sonatest Sales Team: sales@sonatest.com